Nellcor PM100N Patient Monitoring System For Hospitals and Homecare

Product code: PM100NKITFI_PKT

Nellcor PM100N potilasseurantajärjestelmä sairaaloihin ja kotihoitoon

  • Description
  • The Nellcor Bedside SpO2 Patient Monitoring System, PM100N, is intended for prescription use only for continuous monitoring of functional arterial oxygen saturation and pulse rate of adult, pediatric and neonatal patients, who are well or poorly perfused, during motion and no motion conditions. It is intended to be used in hospital, hospital type facilities, and intra-hospital transport as well as in the homecare environments.

    Includes SatSeconds alarm management, a clinician‑controlled feature that can distinguish between real, clinically significant events and transient events by taking into account both the severity and the duration of any desaturation event. Meets IEC 60601-1-11 standards for home health equipment compliant Homecare and Sleep Study modes for safe and effective use of the monitor by lay users and in non‑hospital settings.

    Features

    • Flexible. Affordable. Intuitive
    • Displays real­‑time SpO2 and pulse rate measurements, plethysmographic waveforms and pulse amplitude
    • SatSeconds alarm management
    • Sleep Study mode
    • Homecare mode
    • Adult, Pediatric, Neonate modes
    • Intuitive, easy­‑to­‑read, color, multiple­‑language user interface with on­‑screen help messages
    • Easy­‑to­‑use jog dial interface
    • Compact, portable, durable design with built­‑in handle
    • Variable pitch beep tone for point­‑by­‑point differentiation in SpO2
    • 96-hour trend memory

    Technical Specifications

    Performance

    • Measurement Range SpO2 : 1% to 100%
    • Pulse rate 20 to 250 beats per minute (bpm)
    • Pulse amplitude 0.03% to 20%

    Measurement Accuracy

    • Saturation Adult 70% to 100% ± 2 digits
    • Adult and neonate low sat 60% to 80% ± 3 digits
    • Neonate 70% to 100% ± 2 digits
    • Low perfusion 70% to 100% ± 2 digits
    • Adult and neonate with motion 70% to 100% ± 4 digits

    Pulse rate

    • Adult and neonate 20 to 250 bpm ± 3 digits
    • Low perfusion 20 to 250 bpm ± 3 digits
    • Adult and neonate with motion 20 to 250 bpm ± 5 digits

    Electrical

    Instrument

    • Power requirements 100 to 240 VAC, 50/60 Hz, 45 VA
    • Fuse rating Fast‑acting 2 A 32VAC/DC, Fast‑acting 500 mA 32VAC/50DC

    Battery

    • Type Lithium ion
    • Battery capacity Minimum of five hours using new, fully charged battery with no alarms

    Environmental

    • Operating Temperature
      • Instrument 5 °C to 40 °C (41 °F to 104 °F)
      • Transport/Storage Temperature (in shipping carton) -20 °C to 60 °C (-4 °F to 140 °F)
    • Operating Humidity 15% to 93% noncondensing
    • Operating Altitude -170 m to 4877 m (-557 ft to 16,000 ft)

    Physical Characteristics

    • Weight 1.5 kg (3 lbs)
    • Size 82 H x 255 W x 155 D (mm), (3.23 H x 10.04 W x 6.10 D (in)

    Equipment Compliance

    Standards Compliance

    • IEC 60601-1:2005+A1:2012, EN 60601-1:2006/AC:2010
    • IEC 60601-1:1998 + A1:1991 +A2:1995, EN 60601-1:1990 +A11:1993 +A12:1993 +A13:1996
    • IEC 60601-1-2:2007, EN60601-1-2:2007
    • IEC 60601-1-6:2010, EN 60601-1-6:2010 +A1:2013 
    • IEC 60601-1-8:2006, EN 60601-1-8:2006 +A1:2012 
    • IEC 60601-1-11:2010, EN 60601-1-11:2010
    • ISO 9919:2005, EN ISO 9919:2009 
    • ISO 80601-2-61:2011, EN ISO 80601-2-61:2011
    • CAN/CSA C22.2 No. 601.1 M90 
    • UL 60601-1: 1st edition
    • 802.11 B/G/N WLAN connectivity

    Equipment Classifications

    • Type of protection against electric shock: Class 2 (internally powered)
    • Degree of protection against electric shock: Type BF – Applied part •
    • Mode of operation: Continuous 
    • Electromagnetic compatibility: IEC 60601-1-2:2007 
    • Liquid ingress: IP 22 
    • Degree of safety: Not suitable for use in the presence of flammable anesthetics

    Output 

    • Trend data download via wired or USB for archiving or data analysis

    Display/Indicators

    • Pulse amplitude indicator (eight segments) 
    • Visual indicators: Pulse search, audible alarms silenced or off, interference indicator, battery charging, and SatSeconds alarm management clock, pleth wave form Alarms 
    • SatSeconds alarm management 
    • Audible and Visual alarms for high/low saturation and pulse rate, low battery, sensor off, and sensor disconnect 
    • Setting: Default, institutional and last setting
    •  Alarm system delay: <10 s

    Available Modes

    • Standard – Hospital, hospital‑type facilities, and intra‑hospital transport. 
    • Homecare – Simplified monitoring for use in the home by caregivers 
    • Sleep Study – Muted audible and visual queues to aid sleep studies

    Connectivity 

    • Supports wired and USB trend data export to an external personal computer for archiving or data analysis •

    Product code: PM100NKIT